Services » Radiotherapy services



- Installation and servicing of dosimetry equipment manufactured by
   IBA Dosimetry, Germany


- Training and education of the end-user

Our company is personnel and technically equipped in order to provide the full service training to the end-user for working with dosimetric equipment as well as the education in the procedure of specific dosimetric measurements.


- Dosimetric calculations in filling the system for commissioning
 

Linear Accelerators

Quality Assurance

- QA for linear accelerator
- definition of an unique machine scale and correlation
  of operating limits for linear accelerator
- verification and approving of beam flatness and
  symmetry for all photon and electron energy
- establish and validate dmax for all photon and
  electron energy

Beam Data Measurements

- percentage depth dose data for open fields all
  photon and electron energy
- percentage depth dose data for wedged fields all
  photon energy
- output factors for photon energy
- electron applicator (standard cutout) factors for all
  electron energy
- physical wedge factors for all photon energy
- enhanced dynamic wedge (EDW) factors for all
  photon energy
- beam profiles for all photon and electron energy
- beam diagonal profiles of maximum square field for
  all photon energy
- electronic or/and hard format of all beam data
  measured
- data book with dosimetric tables for hand calculation

Beam Configuration in Pinnacle ADAC

- definition of linear accelerator unit in Pinnacle ADAC
  TPS
- correlate linear accelerator configuration in Pinnacle
  ADAC TPS and 4DiTC administration
- photon beam configuration
- electron beam configuration

Commissioning Validation and Reports

- create a water phantom in Pinnacle ADAC TPS
- analyzing and validating beam data for each photon
  energy
- analyzing and validating beam data for each electron
  energy
- verify plan calculations and validate the plan
  delivery with an ionization chamber in a phantom
- approving beam data for each energy and algorithm
- provide a report of all tests performed

 